2. Adjusting the Off-Center Loading Error: Right - Left
Negative error on the right:
- File the edges of the upper thin sites evenly on the front and back, moving
inwards. Small, inwardly tapering surfaces areas will be produced at the thin
sites „C“
Negative error on the left:
- File the edges of the upper thin sites evenly on the front and/or back, moving
inwards. Small, inwardly tapering surfaces areas will be produced at the thin
sites „D“
dms3ce.eps
dms3de.eps
D
C