Do you have a question about the Sartorius CP622-0CE and is the answer not in the manual?
Maximum Capacity | 620 g |
---|---|
Display | LCD |
Weighing Units | g, kg, oz, lb |
Interface | RS-232 |
Operating Temperature | 10 °C to 40 °C |
Calibration | Internal (automatic or manual) |
Power Supply | AC adapter or battery (optional) |
Explains activating/deactivating BPI mode and managing write-protect settings.
Step-by-step guide to activate the Binary Processor Interface (BPI) mode.
Explains how to adapt the A/D converter for zero point offset using voltmeter or software.
Details how to check and adjust for off-center loading errors.
Instructs on filing thin sites for front-back error correction.
Details filing procedures for back off-center loading error adjustment.
Guides on filing thin sites for right-left error correction.
Explains how to perform external linearity adjustment.
Details using service software for balance linearity adjustment.
Details the procedure for adjusting off-center loading errors in double lever systems.
Illustrates filing positions for correcting off-center loading errors.
Details the external span adjustment procedure.
Explains how to overwrite the internal calibration weight using software.
Provides steps for adjusting off-center loading errors in monolithic systems.
Details the 3-point adjustment procedure for off-center loading errors.
Details the procedure for internal span adjustment.
Provides steps for manual linearity adjustment.
Guides on overwriting internal linearization weights.
Continues listing error codes, explanations, and remedial measures.