34 |
Overwriting the Internal Linearization Weight
The factory setting of the internal linearization weights is highly accurate:
-  If overwriting these weights is necessary, this indicates a mechanical or electrical 
error!
-  The balance/scale must be allowed to warm up (6 hours)!
Note:  There are various ways of overwriting the internal calibration weight on CP balances/
scales.
1.  As described in this manual
2.  or using the Sartorius MC1 Server (Version 4.9 or later)
3.  or with the SARTOCAS Program (Version 1.44 or later) for PCs and laptops.
-  Make sure that the balance/scale operating menu is set to the code „1 9 6 
-internal linearization accessible“ (see page 13-14 „Operating Menu Settings“).
-  If you have not done so already, perform external linearization/adjustment. 
-  Slide the access lock switch to the right. It is accessible via an opening on the 
back of the balance/scale.
-  Turn the balance/scale back on by pressing e and then press the q key 
until the readout shown on the left appears.
- Press the t key to tare the balance/scale; then press q to activate the 
overwrite function.
-  The built-in, motorized internal linearization weights are applied, then removed 
automatically; the weight value is stored in the EEPROM of the processor; and 
the balance/scale returns to the normal weighing mode.
-  After completing the linearization routine, reset the code „1 9 6“ to „1 9 3“ 
and slide the access lock switch to the left (menu locked). Replace the capped 
plug to close the opening.
 