22 |
Adjusting External Linearity
Checking the Linearity
Note:   The linearity must be checked according to the Standard Operating Procedure 
WKD-038.
-  Check the linearity of the balance/scale for the entire weighing range in 4-g 
steps.
-  Compare the values displayed with the tolerance ranges given in the „Table of 
Calibration/Adjustment Data“ on pages 42-43.
-  If the errors exceed the permissible tolerance ranges, the linearity must be 
adjusted.
Adjusting the Linearity 
Caution! You can adjust the linearity on CP balances/scales only using the SARTORIUS MC1 
Server Software (Version 4.9 and later) or with the SARTOCAS program.
-  Activate the BPI mode (see page 15).
-  Adjust the linearity with the SARTORIUS MC1 Server (version 4.9 and later) or 
with the SARTOCAS program for PCs and laptops.
  (Please see the program description for instructions on this automatic proce-
dure!).
Adjusting Linearity