22
External Calibration
Settings:
Calibration/adjustment mode: External calibration/adjustment (menu code 1  9  1)
The weight required for calibration/adjustment is defined in the factory settings 
(see “Specifications").
Step Key (or instruction) Display
1. Turn on the balance/scale, e
0.0 g
if necessary.
2. Tare the balance/scale, t
0.0 g
if necessary.
3. Start calibration/adjustment q
+ 5000.0
Calibration weight is displayed CAL
without a unit symbol.
4. Apply the prompted calibration
5000.0
weight (in this example:  CAL
5000 g).
After calibration, the weight
+ 5000.0 g
unit is shown. CAL
5. Remove the calibration weight.
0.0 g