4
Examples of Data Records:
Printout without Data ID Codes (Menu Code 7.1. 1):
The value currently displayed is printed (weight or calculated value with unit)
+ 1530.0 g Weight in grams
+ 58.562 ozt Weight in Troy ounces
+ 253 pcs Piece count
+ 88.2 % Percentage
+ 105.8 o Calculated value
Printout with Data ID Codes (Menu Code 7.1. 2):
The current value displayed can be printed with a data ID code of up to 6 characters at the
beginning of a line. You can use this data ID code, e.g., to designate a weight readout as
a net weight (N) or a calculated value as a quantity (Qnt).
N + 153.0 g Current net weight
N1 + 153.0 g Current net weight (with data in 2nd tare memory)
T1 + 10.2 g Value in 2nd tare memory
Qnt + 253 pcs Calculated quantity (piece count)
Prc + 88.2 % Calculated percentage
Res + 153.0 g Calculated result
Print Application Parameters (Menu Code 7.1. 2):
You can generate a printout of one or more of the values configured for initialization of an
application as soon as you initialize the balance/scale.
wRef + 1.432 g Counting: average piece weight
Wxx% + 120.12 g Weighing in percent: reference weight for the
selected percentage
Auto Print (Menu Code 6 .1. 3 or 6 .1. 4):
You can have the weight readout printed automatically. The display update interval depends
on the operating status of the balance/scale and on the balance/scale model.
N + 153.0 g Net weight
Stat Display blank
Stat L Display underload
Stat H Display overload