Density Determination
Display symbol: W
Purpose
This application program lets you determine the density of solid substances using the buoy-
ancy method. You can have results displayed with one decimal place, or no decimal places:
see “Configuration.”
Note: the sample holder and suspension wire used in the example below are not included
with the balance/scale.
Parameter settings:
Applic.: Density: Dec.Plcs: 1 Dec. Pl. (menu code 2. 9. 1. 2)
Example: Determining the density of a solid sample.
Step Key (or instruction) Display/Data output
1.
Attach sample holder to suspension wire
2. Tare the balance/scale w 0.0 g
3. Start application program
u
4. Confirm “Airdisplay u Air
?
5. Determine the weight
of the sample in air:
Place sample on the balance/scale +
20.0 g
*
6. Store value for weight in air u
7. Remove sample from the balance/scale water
?
8. Determine weight in liquid:
place sample in holder
9. Confirm “waterdisplay
u 0.0 g
*
10. Immerse sample in liquid + 15.0 g
*
11. Store value for weight in u + 4.0
o
*
liquid, view result, and print Wa + 20.0 g
Wfl + 15.0 g
Rho 4.0 o
12. Delete result c
13. Repeat as desired, starting from Step 3.