Interface Port
55
Purpose
Your balance is equipped with an interface
port for connection to a computer or other
peripheral device.
You can connect a computer to change,
start and/or monitor the functions of the
balance and the application programs. 
Features
Type of interface:  Serial interface
Operating mode:  Full duplex
Standard:  RS-232
Transmission rates:  
150, 300, 600, 1200, 2400, 4800, 9600
and 19,200 baud
Parity:  Mark, space, odd, even
Character format:  
1 start bit, 7-bit ASCII, parity, 
1 or 2 stop bits
Handshake:  
2-wire interface:  
via software (XON/XOFF)
4-wire interface:  
Hardware via handshake lines (CTS/DTR)
Operating mode:  SBI
Data output format of the balance:  
16 or 22 characters
Factory settings:
Transmission rate:  1200 baud (5 1 4)
Parity:  Odd (5 2 3)
Stop bits:  1 stop bit (5 3 1)
Handshake:  Hardware, 2 characters 
after CTS (5 4 2)
Operating mode:
Standard Sartorius interface SBI (5 5 1)
Print manually/automatically:  
Manual after stability (6 1 2)
Preparation
§ see “Pin Assignments" and 
“Pin Assignment Chart"
Identification of Non-Verified Digits
Non-verified digits when “e#d” 
are automatically identified on the
printout: 
Select universal printer: 
menu code 5 5 2. 
Brackets are used to identify 
non-verified digits.