Maximum Capacity | 1300 g |
---|---|
Readability | 0.01 g |
Calibration | External |
Display | LCD |
Repeatability (Std. Dev.) | 0.01 g |
Linearity | ±0.02 g |
Stabilization Time (typical) | 2 seconds |
Power Supply | AC Adapter |
Weighing Units | g, ct, oz, dwt |
Interface | RS232 |
Operating Temperature | 10°C to 30°C |
Weight | 2.5kg |
Controls access to calibration and adjustment functions, including auto zero.
Restores the balance to its original factory settings and resets SBI mode.
Explains the process of adjusting the zero-point offset using a DVM and specific components.
Describes how to test for off-center loading errors by placing weights in different positions.
Explains OIML guidelines for off-center loading tests and placement of test weights.
Details the method for adjusting off-center errors by filing thin parts on the load receptor.
Details how to test off-center errors for different pan types using specific models.
Explains how to perform span adjustment using the operating menu and calibration weights.
Outlines the process for checking linearity using specific weighing steps and tolerances.
Describes using service software to adjust the balance linearity after checking.
Guides on performing repeatability tests using specified weights and calculating the standard deviation.
Explains the 3-point measurement and adjustment process for double-lever systems.
Details the 3-point measurement method to determine and adjust off-center loading errors.
Describes adjusting off-center errors by filing thin parts on specific models.
Details the process of replacing the main PCB, including data backup and reprogramming.