EasyManua.ls Logo

Sartorius Gem GE1302 - Replacing the PCB; PCB Replacement Procedure

Sartorius Gem GE1302
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
Aut12628.jpg
E
D
F
AUT12754a.JPG
AUT12423.JPG
G
H
Replacing the PCB
In the case of defects, do not attempt to make any repairs at the component
level, but replace the entire subassembly.
Note: Before replacing the main PCB, use the service software to read out
and store the balance/scale’s data record, if possible. If this is not
possible, order a pre-programmed PCB, indicating the balance/
scale model and serial number.
Open the balance/scale (see page 48).
Disconnect all cables.
Remove the fastening screws (G) from the main PCB.
Remove the old PCB
Take the zero-point offset resistor from the old PCB and place it on the new
PCB
Follow the above instructions in reverse order to install the new main PCB
The balance/scale must now be programmed with the required data record
(unless you have installed a pre-programmed PCB)
The following values must be checked and, if necessary, adjusted:
Zero-point offset
Linearity
Span

Related product manuals