Readability | 0.001 g |
---|---|
Display | LCD |
Repeatability | 0.001 g |
Linearity | ±0.002 g |
Calibration | External |
Power Supply | AC adapter |
Interface | RS232 |
Steps for enabling BPI mode and disabling write protection for service.
Steps for connecting the BPI adapter to activate the BPI mode.
Final steps for connecting equipment and verifying BPI mode activation.
Procedure for checking and adjusting the overload stop using service software.
Step-by-step guide for checking and adjusting the overload stop.
Procedure for adjusting the zero-point offset voltage using a DVM.
Method for calculating and comparing repeatability results against tolerances.
Procedure for checking off-center loading error at different positions.
Steps for adjusting span using operating menu for specific models.
Procedure for checking and adjusting linearity for double lever systems.
Steps for adjusting off-center loading error using adjustment screws (P).
Using a 3-point measurement to determine and adjust off-center loading error.
Adjusting off-center error by filing thin places on the upper guide.
Steps for reading data, removing, and installing a new main PCB.
Causes and solutions for common error codes displayed on the balance.