Sartorius
G 150 D, G 200 D, G 500
Electronic Gem Scales
Installation and Operating Instructions
98648-002-26

Questions and Answers

  • S
    sandra37Aug 3, 2025
    Why does the weight readout change constantly on my Sartorius Scales?
    • J
      James GonzalesAug 3, 2025
      The weight readout might change constantly on your Sartorius Scales due to an unstable environment, excessive vibration, a draft, or a sample that does not have a stable weight. To resolve this, try setting up the scale in another area to avoid vibration or drafts. You can also access the menu to adapt the scale to your specific weighing environment.
  • N
    nicholas75Aug 7, 2025
    What to do if Sartorius G 150 D weight display shows “H”?
    • M
      matthewgrantAug 7, 2025
      If the weight display on your Sartorius Scales shows “H”, it means the load exceeds the scale's capacity. Unload the scale to resolve this issue.
  • M
    Mario BellAug 11, 2025
    Why is the weight readout obviously wrong on my Sartorius Scales?
    • T
      twilliamsAug 11, 2025
      If the weight readout on your Sartorius Scales is obviously wrong, it could be due to several reasons: the scale hasn't been calibrated, it hasn't been tared before weighing, or the air bubble of the level indicator is not within the circle. To fix this, calibrate the scale, tare it before weighing, and ensure the scale is leveled properly.
  • I
    ifreemanAug 15, 2025
    How to fix the “CC“ code on my Sartorius Scales display?
    • R
      robertellisAug 16, 2025
      If the code “CC“ remains on the display of your Sartorius Scales, it indicates that the scale is either not ready to calibrate or is in the warmup phase, or the weighing system is affected by draft or vibration. Ensure the scale warms up for at least 30 minutes after plugging it in. If the issue persists, adjust the appropriate code by accessing the operating program's menu to address any draft or vibration issues.
  • B
    Bradley GarciaAug 20, 2025
    What to do if Sartorius Scales weight display shows “CE”?
    • M
      Michael SchroederAug 20, 2025
      If the weight display on your Sartorius Scales shows “CE”, it means that a zero readout was not indicated when the CAL key was pressed, or the scale is loaded. Press the tare control and re-press the CAL key to resolve this.
  • E
    Erin PachecoAug 25, 2025
    What to do if Sartorius G 150 D Scales weight display shows “L” or “CH2”?
    • K
      katherinebraunAug 25, 2025
      If the weight display on your Sartorius Scales shows “L” or “CH2”, it indicates that the pan support and/or weighing pan is not in place. Ensure that the pan support and the weighing pan are correctly placed on the scale.
  • L
    Laura RiceAug 28, 2025
    How to fix if no segments are indicated in the weight display of Sartorius G 150 D Scales?
    • J
      Jenna ThomasAug 29, 2025
      If no segments are indicated in the weight display of your Sartorius Scales, it could be due to a lack of AC power or the AC adapter not being plugged in. Check the AC power supply and ensure the AC adapter is properly plugged in.

Summary

About the Product

Storage and Shipping Conditions

Equipment Supplied

Installation Instructions

Ambient Conditions

Choose a suitable place to set up your scale, avoiding heat, drafts, vibrations, and chemical atmospheres.

Connecting Electronic Devices (Peripherals)

Unplug AC adapter before connecting/disconnecting peripherals to the scale's interface port.

Startup

Mounting the Draft Shield

Place the draft shield on the scale, use a screwdriver to mount it, and insert the base plate.

Connecting the Scale to AC Power

Ensure the AC adapter voltage matches local rating; use only original Sartorius adapters.

Power and Safety Setup

Voltage Selection

Select voltage only for portable power supply with European-type plug (rounded prongs).

Safety Precautions

Class 2 power supply/adapter can be plugged in without extra precautions; interface is electrically connected to scale housing.

Leveling the Scale

Leveling the Scale Using the Level Indicator

Level the scale using feet (15) so the air bubble is centered in the level indicator (9).

Operating the Scale

Display On;Off and Status Messages

Use ON/OFF key or tare control for display; scale ready after standby. Status messages explained.

Automatic Self-Test

Electronic circuitry test runs automatically after power-on, showing a specific final readout.

Taring Operation

Zeroing the Display

Press tare control to zero display before weighing if current reading is not zero.

Weighing Process

Selecting Range and Weighing

Select range (W1/W2), use accessory, read display when stability symbol appears.

Calibration

Scale Operating Menu

Accessing the Menu

Hold tare control, press ON/OFF key briefly when display is off; 'CH5' appears.

Reading Menu Settings (List Mode)

Press tare control for numeral; 'L' indicates list mode (read only).

Changing Menu Settings

Change Mode Operation

Slide switch (7) to 'C' mode; press tare control for numeral changes. 'o' shows current setting.

Returning to Weighing Mode

Slide switch (7) back; 'L' displayed. Press tare control for '0' to return.

Auto Zero Function

Automatic zero tracking function (can be turned off via menu code).

List of Programmable Menu Codes

Troubleshooting Guide

Care and Maintenance

Cleaning Procedures

Use a mild detergent wet cloth. Avoid liquid entry. Wipe dry with a soft cloth.

Safety Inspection Guidelines

If safe operation is uncertain, turn off, disconnect power, notify service. Only authorized techs can repair.

Accessories (Options)

Interfacing Devices with the Scale

Specifications

Sartorius G 150 D Specifications

General IconGeneral
Maximum Capacity150 g
Readability0.1 mg
DisplayLCD
Linearity±0.2 mg
Repeatability0.1 mg
InterfaceRS232
Operating Temperature+10°C to +30°C