Calibration
Calibration must be done independently for each weighing range
(
W1/W2
).
To calibrate, unload the scale and tare.
Press the CAL key
(5)
once the display indicates a zero readout.
The Status symbol "
C
" will now be indicated in the weight display.
lf "
CE
" is displayed instead, zero the display and press the CAL key
again.
After a few seconds,
"CC"
will be displayed, followed by a zero
readout along with the symbol of the particular weight unit selected.
Anacoustic Signal will indicate the end of the calibratiön process.
You can block access to the calibration function - refer to the "Scale
Operating Menu."
(This calibration function is accessible whenever the scale operating
menu is in the
"C"
[Change] mode.)