EasyManua.ls Logo

Sartorius G 150 D - Specifications

Sartorius G 150 D
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
Specifications
Model
G 150 D
Weighing range structure DualRange
Weighing range ct/g 152/111
Readability ct/g 0.001/0.01
starting with serial no. 38050001
0.001/0.002 g
Reproducibility
(standard deviation)
ct/g
±0.001/0.01
starting with serial no. 38050001
±0.001/0.002 g
Linearity ct/g
±0.002/0.02
starting with serial no. 38050001
±0.002/0.006 g
Stabilization time (average) s 2
Display update s 0.1 – 0.8 (selectable)
Adaption to ambient conditions
and application requirements
By selection of one of 4 optimized filter levels
Stability range d 0,25 ... 64 (selectable)
Ambient temperature range K 273 ... 313 (0
o
C ... +40
o
C)
Moisture-proof rating according
to DIN 40040
Class F, non-condensing
Sensitivity drift within 283 … 303 K
/K
±410
-6
Weighing pan size mm Ø63
Balance housing (W x D x H) mm 185 x 200 x 210
Net weight, approx. kg 3.5
Power requirements
(voltage + frequency)
115 or 230 V depending on the AC adapter used
50 – 60 Hz
Allowable voltage fluctuation -15% ... +10%
Power consumption VA 7 (average.)
Interface (optional RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;
transmissions rate: 150 ... 9600 Baud