15
List of the Programmable menu Code Settings
C 1 3 1
Page Line Word
Menu of the Balance Operating Program
(active parameters)
Code
Ambient Conditions
Code
Internal Calibration
C 1 1 1
Very stable
C 1 7 1
Accessible
C 1 1 2
Stable (W2)
C 1 7 2
Access blocked
C 1 1 3
Unstable (W1)
C 1 1 4
Very unstable
Special Information
Code
Program Lock
Code
Stability Range
C 4 1 1
OFF
C 1 2 1
0.25 digit
C 4 1 2
ON
C
1
2
2 2
0.5 digit
C 1 2 3
1 digit
Code
Acoustic Signal
C 1 2 4
2 digits
C 4 3 1
ON
C 1 2 5
4 digits
C 4 3 2
OFF
C 1 2 6
8 digits
C 1 2 7
16 digits
Code
Weight Units
1)
C 1 2 8
32 digits
c 5 1 1
Grams g (W2)
2
C 1 2 9
64 digits
c 5 1 2
Kilograms kg
c 5 1 3
Carats ct (W1)
Code
Display Format
c 5 1 4
Pounds lb
C 1 3 1
Last decimal ON
c 5 1 5
Ounces oz
C 1 3 2
Last decimal OFF
c 5 1 6
Troy-Ounces ozt
C 1 3 3
Last decimal at stability
c 5 1 7
Parts/Pound o
C 1 3 4
All decimals at stability
c 5 2 1
Hongkong taels tl
c 5 2 2
Singapore taels tl
Code
Tare Parameter
C 5 2 3
Taiwanese taels tl
C 1 4 1
Without stability
C 5 2 4
Grains gr
C 1 4 2
At stability
C 5 2 5
Pennyweights dwt
C 5 2 6
Mommes o
Code
Auto Zero
C 5 2 7
Milligrams o
C 1 5 1
ON
C 5 2 8
Austrian carats o
C 1 5 2
OFF
C 5 2 0
Call Program Line
Code
External Calibration
C 5 0
Call program Page
C 1 6 1
Accessible
C 1 6 2
Access blocked
C 0
End of Programming
Additional Parameters for the data Output format at the interface port and for calculation programs are avail-
able on request.-Please refer to the “Accessories."
1)
You can choose any weight unit as long as it can be displayed in the particular weighing range you
selected (for example do not set the code for "kg" when you are using a 0.1 mg balance).
2)
not for the G 200 D/ G 150 D