Do you have a question about the Sartorius Research RC 210S and is the answer not in the manual?
Requirements for working on Sartorius high-resolution balances.
Notes on the optimum ambient conditions for balance operation.
Information on essential measuring and test equipment for balance checks.
Critical safety advice regarding unsuitable measuring and test equipment.
Lists special tools required for Sartorius Research Balances.
Lists related documentation for RC/MC balances.
Explains the function of various keys on MC and RC balance models.
Step-by-step guide to setting identification codes, time, and date.
Details on entering an 8-digit ID code for compliant records.
Instructions for setting time in 12-hour or 24-hour format.
Procedure for setting the day, month, and year.
How to view the balance's hardware and software version numbers.
Conditions and steps before checking balance specifications.
Requirements for the balance installation environment.
Initial steps for balance setup and warm-up.
Guidelines for selecting and using calibrated test weights.
Overview of the testing procedure and documentation.
Procedure for testing balance reproducibility using specific weights.
List of weights for reproducibility testing.
Step-by-step weighing process for reproducibility.
Method for calculating and evaluating reproducibility results.
Formula and method for calculating standard deviation.
Procedure to test and adjust balance performance with off-center loads.
Weights used for off-center load testing.
Steps for conducting the off-center loading test.
How to evaluate results for off-center loading tolerances.
Specific weights for internal span adjustment.
Instructions for recording tolerances and conventional mass.
Calculation for determining the conventional mass of test weights.
Calculation for the sum of tolerance and uncertainty.
Steps for performing the internal span adjustment weighing.
How to calculate and evaluate test results for span adjustment.
Conditions and actions if linearity test fails.
Weights for linearity testing.
Tare weights for linearity testing.
Step-by-step process for checking balance linearity.
Order of adjustments: reproducibility, off-center load, linearity, span.
Steps before adjusting, including service prerequisites and BPI mode.
Coarse adjustment procedure for off-center load.
Fine adjustment procedure for off-center load.
Reference to the linearity check procedure on page 12.
Methods for external linearity adjustment using software.
Procedure for internal linearity adjustment using balance menu.
Instructions for safely opening the balance housing.
Steps for removing printed circuit boards and enclosure parts.
Detailed procedure for removing the draft shield components.
Procedure for cleaning the weighing system components.
Steps to access the weighing system for cleaning.
How to inspect and clean the weighing system gap and base plate.
Steps to reassemble the weighing system after cleaning.
Steps to expose the calibration weight system for cleaning.
Detailed instructions for replacing the display PCB.
Lists common software error codes, meanings, and solutions.
Lists hardware error codes, meanings, and solutions.