Gem
plus
Series
Model GC1603S-0CE GC803S-0CE
Type BC BL 100 BC BL 100
Accuracy class
1
) kk
Maximum capacity, Max.
1
) ct 1,600 800
Readability ct 0.001 0.001
Tare range (subtractive) ct 100% of the maximum capacity
Verification scale interval, e
1
) mct 10 10
Minimum capacity, Min.
1
) ct 0.1 0.1
Stabilization time (average) s 2
Range of use ct 0.1–1,600 0.1–800
according to CD
1
)
Allowable ambient
operating temperature °C +15… +25 (50°F to 77°F)
External calibration weight g 200 + 100 (E2)
(of at least accuracy class) 100 (E2)
Net weight, approx. kg 6.1
Pan size (inner diameter)* mm 80 80 ‡
Pan area* cm
2
64 64
Weighing chamber height
(from pan to sliding cover) mm 162
Dimensions (WƒDƒH) mm 213ƒ342ƒ270
AC power source/ V~ AC adapter STNG6 230 VAC or 115 VAC,
Power requirements +15%…–20% (protection rating IP20)
Frequency Hz 48 – 60
Power consumption (average) VA maximum: 16; average: 8
Approx. hours of operation
with the YRB05Z
rechargeable battery pack h 22 22
Selectable weight units Grams, milligrams, carats
Built-in interface RS-232C-S/V24-V28; 7-bit;
parity: even, mark, odd, space;
transmission rates 150...19,200 baud;
1 or 2 stop bits; software/hardware handshake
1
) CD = Council Directive 90/384/EEC on non-automatic weighing instruments;
applicable to the European Economic Area
* Three-sided weighing pan: ‡ = Diameter of inner circle.
The cross-hatched section can be fully utilized.