24
External Calibration
Settings:
Calibration/adjustment mode: External calibration/adjustment (menu code 1  9  1)
The weight required for calibration/adjustment is defined in the factory settings 
(see “Specifications").
Step Key (or instruction) Display
1. Zero the balance. (TARE)
0.0 g
2. Start calibration. (CAL)
+ 5000.0 g
CAL G
3. Apply the prompted calibration 
weight (in this example, 5000 g).
4. The balance is calibrated
– 0.2 g
(displayed only if menu  CAL ± G
code 1 10 2 is set).
5. If the “Calibrate, then auto 
Adjust
*
adjust" setting is selected  CAL
(1  10  1), the balance is now 
adjusted automatically.
6. The calibration/adjustment  
CC
sequence is completed. CAL
7. After calibration/adjustment, the
+ 5000.0 g
weight is displayed with weight unit.
8. Remove the calibration weight.
0.0 g
* Adjust shown only if menu code 1 10  2 is set.
Important note: Afterwards, do not perform internal calibration/adjustment again.