Setting IDs, Time, Date
and Display Brightness
Purpose
Configuring measurement environment parameters
for ISO/GLP-compliant data records.
Setting date and time (for ISO/GLP-compliant
records only).
Adapting display to ambient lighting conditions.
Features
Enter up to 8 characters to identify a measurement
series. Permissible characters include the numbers
0 through 9 and the dash or minus sign (“–").
A dash is output as a space on printouts. Leading
zeroes are not output.
Date and time at beginning and end of ISO/GLP-
printouts.
Display brightness 1):
0 = off; levels of brightness: 1 through 9
Key functions during configuration:
Activate input of IDs, time and date:
Switch the balance off and then on again by
pressing (E); while all segments are displayed,
press the (F) key briefly
Scroll upward : Press (CAL)
Scroll to the right : Press (A)
Confirm input and toggle between IDs,
time and date: Press (TARE)
Save settings and exit menu:
Press and hold (TARE) (> 2 sec.)
1
) No display backlighting on models CPA2P…,
CPA26P(-0CE) and CPA225D(-0CE)
32