44
Selecting the Calibration/
Adjustment Parameter
The setting
Selection mode
must be selected in the Setup menu under
“Balance functions: Calibration/adjustment:
CAL key function” (factory setting).
After pressing the
Cal soft key, you can
choose from among the following settings
by pressing the
Select soft key:
– External calibration/adjustment with
the preset calibration weight:
Ext. cal./adj.;
factory-def. wt.
– External calibration/adjustment with
a calibration weight determined by the
user:
Ext. cal./adj.;
userdefined wt.
– Internal calibration/adjustment
Internal cal./
adjustment
– Reproducibility test
reproTEST
G Start the desired routine:
Press the
Cal soft key again
In the selection mode: Perform external calibration followed by automatic adjustment with the factory-set weight
Configuration:
factory settings
Step Key(s) (or instruction) Display/Output
1. Select Calibration
Cal soft key
2. Select external calibration/adjust- 2 +
Select soft key
ment with factory-defined weight
3. Start external calibration/
Start soft key
adjustment
4. Place the weight on the Place weight on balance
balance (e.g., 2,000.00 g)
Minus sign –: Weight too low
Plus sign + Weight too high
no plus/minus sign: Weight o.k.
This is displayed after calibration,
for approx.10 seconds:
(on verified balances, the difference between
the displayed weight and the true weight (mass)
is displayed)
After adjustment, the following is displayed:
5. Unload the balance
(ISO/GLP/GMP printout: see page 116)
CAL: Extern. adj. factory-def. wt.
CAL: Extern. adj. factory-def. wt.
CAL: Extern. adj. factory-def. wt.