158
Verified Models with EC Type Approval
Model ME36S-0CE ME5-0CE SE2-0CE
Type KD BM KC BN KC BN
Accuracy class* kkk
Scale interval d* µg 1 1 0.1
Max. weighing capacity* g 31 5.1 2.1
Verifcation scale interval e* mg 1 1 1
Min. capacity* mg 0.1 0.1 0.01
Tare range (subtractive) <100% of the max. weighing capacity
Application range* mg 0.1-31,000 0.1-5,100 0.01-2,100
Response time (average) s 14–18 <10 <10
Allowable operating temperature range 273 … 313 K (0 … +40°C, 32°F … 104°F) with “isoCAL” function
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate (depends on filter level) 0.2 – 0.4
Pan size mm d 30 d 30 d 20
Balance dimensions (W + D + H)
– Weigh cell mm 223 + 410 + 232 122 + 316 + 122 122 + 316 + 122
– Display and control unit mm 254 + 320 + 106 254 + 320 + 106 254 + 320 + 106
Net weight, approx.
– Weigh cell kg 9.8 3.3 3.3
– Display and control unit kg 3.5 3.5 3.5
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 – 60 Hz
Power consumption (AC adapter) max. 23 VA
Power consumption, weigh cell and approx. 7 W (typical)
display and control unit only
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
DC nominal supply voltage 10.5 to 25 Vdc
Selectable weight units Grams and milligrams
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
* = according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
Specifications