Operating the Balance
48
Below-Balance Weighing
A below-balance weighing hanger is located on the bottom of the balance.
Not allowed in legal metrology applications
G Open cover plate (1) on the bottom of the balance
G Lift off weighing pan 2
G Unscrew hanger 3
G Insert opposite end of hanger 3 into port and refasten
G Place weighing pan 2 back on balance
G Hang sample on the notched hook
§ Remove both screws from beneath the weigh cell and detach the cover
§ Attach a wire to the sample, if necessary, and hang it on the notched hook
$ If necessary, install a shield for protection against drafts
Preparation for Operation
G Turn on the balance: press e
> The Sartorius logo is displayed
$ If it is time for the next maintenance, the following appears:
To exit this screen: press the
oo soft key
$ Call your nearest Sartorius Service Center to schedule a maintenance appointment
$ To tare the balance, if desired, press w
> The S symbol is displayed when a verified balance is zeroed or tared (±0.25 digit)