Example
Document the evaporated amount of a sample with defined surface, temperature and air pressure at preset intervals of 1 minute, 30 seconds.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 2: Time-controlled functions
Setup: Balance/scale functions: Taring: Without stability
Setup: Printout: Application-defined output: Stability parameter: Without stability
Step Press key(s) (or follow instructions) Display/Output
1. Turn on the balance and configure e
the settings as indicated above
2. Delete stored values, if necessary c
3. Place container with sample t
on the balance and tare
4. Enter time interval: 1 minute, 1 . 3 0
30 seconds
5. Store time interval
Interv soft key
6. Begin documentation
Start soft key
(Time remaining until the
next printout is displayed
in the text line)
Printout of evaporated amount
every 1
1
2 minutes
7. Stop the documentation procedure Stop soft key
Time: 15:19:50
N - 0.37158 g
Time: 15:21:20
N - 0.33215 g
Time: 15:22:50
N - 0.30187 g
Time: 15:24:20
N - 0.40518 g