Balance Functions (Overview)
ο factory setting
user-defined setting(s)
Setup Calibration | Cal | isoTST External calibration | adjustment** with
Balance Adjustment key function factory-defined calibration weight
functions External calibration | adjustment**;
user-defined weight
Internal calibration | adjustment
Internal linearization
Cal key blocked
reproTEST
ο Selection mode for cal. | adjustment
Calibration | ο Calibrate, then auto adjust
adjustment sequence Calibrate, then manual adjust
isoCAL function Off
Only adjustment prompt
isoCAL and linearization on
On and reset application
ο On without resetting application
Start automatic User-defined adjustment Enter adjustment time 1: (hh.mm)
adjustment times Enter adjustment time 2: (hh.mm)
Enter adjustment time 3: (hh.mm)
ο isoCAL
Print GLP ο Automatic if GLP is selected
adjustment record On request,
from record memory
Parameter for Weight set ID (W ID): Enter 14 characters max.
external weight Calibration |
adjustment weight: Enter exact weight
Adapt filter Minimum vibration
(ambient ο Normal vibration
conditions) Strong vibration
Extreme vibration
Application filter ο Final readout
Filling mode
Stability range 1⁄4 digit
1⁄2 digit
1 digit
ο 2 digits
4 digits
8 digits*
Taring* Without stability
ο After stability
Auto zero ο On
Off
* = not applicable to verified balances
** = only external calibration is possible for verified balances