Loading...
Printout Parameters (Overview)
ο factory setting
user-defined setting(s)
SETUP Application- Stability parameter Without stability
1)
Printout defined output o With stability
Print on request, then tare o Off
On
Auto print upon Off
initialization o All values
(of the application) Only main values
Configured printout See own chapter starting
on the next page
Automatic Stability parameter o Without stability
output At stability
of displayed Automatic after
value weight change
2
)
Stop auto print Use print key p
/r
o Not possible
Time-dependent o 1 display update
auto print 2 display updates
10 display updates *
100 display updates *
Output to Serial communication Off
interface ports (PERIPHERALS) o Application-defined output
Automatic output of displayed value
Serial printer) Off
(PRINTER) o Application-defined output
Line format For raw data (16 characters)
o For other apps/
GLP (22 characters)
ISO/GLP o Off
printout Only for calibration/adjustment
Always on
Identification # Lot ID (L ID): Enter 20 characters max.
ID1: ID1
ID2: ID2
ID3: ID3
ID4: ID4
Factory settings: No
printout only Yes
* = changing settings not applicable to verified balances
1)
= Information on use in legal metrology: Only permitted for control purposes; printouts are not allowed
2)
= auto print when load change is > 10 d and stability is reached: no printout until residual difference in load value is < 5 d