EasyManua.ls Logo

Sartorius ME5-0CE - Preparation for Differential Weighing; Steps for Preparing Differential Weighing Application; Differential Weighing Parameter Configuration

Sartorius ME5-0CE
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Printout for Differential Weighing
Generating Printouts
Automatically
The configured backweighing printout
is generated automatically after
backweighing, if one of the following
settings is selected Setup: Application 1:
Differential weighing: Generate print-
out:
Automatic after
backweighing
Auto after init.weigh
+ backweigh
Auto after tare-,
init- +backweigh.
Generating Printouts Manually
The individual printout is generated
when the p/r key is pressed
while there is a tare, initial or back-
weight on the balance, or when d
is pressed to toggle applications.
You can generate the configured
printout manually after backweighing
if you press the p/r key while
the display page for the results is
shown.
To generate the statistics printout,
press the p/r key
when the display page for statistics
is shown,
when the samples with a desired number
of backweighing operations is selected
(for example, statistics on all samples
with 2 backweighing operations).
77
The following printout is generated:
Backweighing Printout
(Example)
-------------------- Dotted line
16.11.1999 14:55:12 Date/time
Lot CH12345 Lot ID
Sample 14 Sample number
ID CX88 Sample ID
T1 + 23.45821 g Tare weighing (with PT1 selected)
N1 +125.57234 g Initial weight
R (3)+103.68442 g Backweight (residue as weight)
R + 82.57 % Residue in percent
D - 21.88792 g Loss as a weight
D - 17.43 % Loss in percent
Fact + 1.10345 Calculation factor
D-Res - 24.15 o Calculated loss
Ratio1+ 21.11 % Ratio 1
Ratio2+ 121.11 % Ratio 2
-------------------- Dotted line

Table of Contents