Sartorius MSA6.6S - (Depending on the Set Filter Level) 0.1 - 0.4 0.1; Display Result; Protected against Dust and Water; Weighing Chamber Height

Sartorius MSA6.6S
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cubis MSA User Manual 179
Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models MSA5203S-0CE MSA5203P-0CE MSA3203S-0CE MSA3203P-0CE
Accuracy class* k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/2/5 1 1/10
Weighing capacity max* g 5200 1200/2400/5200 3200 1010/3200
Calibration value e* mg 10 10 10 10
Min. load min* mg 100 100 100 100
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.1 – 5200 0.1 – 5200 0.1 – 3200 0.1 – 3200
Min. initial weight** g 1.5 1.5 1.5 1.5
Typical stabilization time s 1 1 1 1
Typical measurement time s 2 2 2 1.5
External standard calibration value g 5000 (E2 5000 (E2 2000 (E2 2000 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 ... +40 °C
Without “isoCAL” function: +15 ... +25 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
Models MSA2203S-0CE MSA2203P-0CE MSA1203S-0CE
Accuracy class* k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/10 1
Weighing capacity max* g 2200 1010/2200 1200
Calibration value e* mg 10 10 10
Min. load min* mg 100 100 100
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.1 – 2200 0.1 – 2200 0.1 – 1200
Min. initial weight** g 1.5 1.5 1.5
Typical stabilization time s 1 1 1
Typical measurement time s 1.5 1.5 1.5
External standard calibration value g 2000 (E2) 1000 (E2) 1000 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 ... +40 °C
Without “isoCAL” function: +15 ... +25 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26

Table of Contents