Setting “TEXTS” in the Display, “LONG” or “SHORT”
Either short or long display prompts for operator guidance can be shown.
Level 1 Level 2 Level 3 Level 4
SETUP § Press the L key [ENTER]
EXTRAS § RS keys: select “EXTRAS”
§ Press the L key [ENTER]
TEXTS § RS keys: select “TEXTS”
§ Press the L key [ENTER]
LONG § RS keys: select “SHORT”,
o SHORT confirm by pressing the L key.
§ Press the c key (Clear) several times
to exit the menu
Resetting the Scale: “RESET”
If necessary, you can reset the scale to factory settings.
Note:
If a password was activated, the correct password must first be entered.
Level 1 Level 2 Level 3 Level 4
SETUP § Press the L key [ENTER]
RESET § RS keys: select “RESET”
§ Press the L key [ENTER]
MENU § RS keys: select “MENU”.
§ Press the L key [ENTER]
YES § Use the RS keys to select “YES”
o NO § Press the L key [ENTER]; “o” will appear:
the new code is set
§ Press the c key (Clear) several times
to exit the menu
Setting Codes
Under the setting “CODES,” the menu items are displayed in code 1.1.1.1.
Level 1 Level 2 Level 3 Level 4
LANGUAGE § R key: select “LANGUAGE”
§ Press the L key [ENTER]
GERMAN § RS keys: select “CODES”
etc. § Press the L key [ENTER]; “o” will appear:
the new code is set
o CODES § Press the c key (Clear) several times
to exit the menu.
Note:
To obtain a detailed list of the menu codes, please ask your nearest Sartorius office.