Undoing All Menu Code Changes – Reset Function
This function enables you to reset all menu codes back to the original factory
settings, which are identified by an “*” in this manual. This can be very useful,
for instance when you are not sure what changes have been made in the
code settings.
Reset function Code Changes
On (active) 9––1
Off 9––2 *
Connecting the YDP01IS-0CE Printer to a QC Scale
You can connect a QC scale to a YDP01IS-0CE industrial printer rather than the
YDP03-0CE printer via the cable supplied. You should make the following
changes in the code settings of your QC scale:
Function Code Changes
9,600 baud 5 1 7
Space parity 5 2 2
2 stop bits 5 3 2
Handshake: 2 characters 5 4 2
Printout product no.: ID and line feed 7 4 4 *
Note:
The date/time printout is available only with the YDP03-0CE printer.
You can also print labels on the QC scale with the YDP01IS-0CE printer
(without bar code printout).
*= factory setting