Counting
Counting u/w key Code 2 1 4*
Symbol displayed: P
The counting program allows automatic conversion of weights into piece counts
based on a reference sample quantity and weight. A weight readout is stored as
a reference sample quantity (factory setting: 10 pcs = pieces).
General Functions in the Counting Program
Function/Application Entry Mode User Data
Guide Output
Display ID
Clear the counting c/c key + CF –
application function u/w key
Display information: i/i key + reF –
reference sample u/w key
quantity “nRef”or average
piece weight “wRef”
Data output: i/i key + nRef
reference sample quantity p/p key + reF wRef
“nRef” or average u/w key
piece weight “wRef”
With menu setting 7 1 2:
automatic!
Entry of an average Numeric keys 0 –9
piece weight “wRef” and . + u/w key – –
Entry of a reference sample Numeric keys 0–9
quantity “nRef” without . + ––
u/w key
A number entered manually without a decimal point is always a piece count!
A number entered manually with a decimal point is always a weight value!
* = factory setting