There are 3 different ways you can then re-start the counting process:
1. Press u/w again. The weighed value for calculation of the reference
weight will now be stored with a longer weighing time and “increased stability.”
2. Empty the container and place it on the scale again. Press =/z
or o/t. The tare value is now stored with a longer weighing time and
“increased stability.” Place the pieces back in the container and press
u/w. The weighed value for calculation of the reference weight will now
be stored with a longer weighing time and “increased stability.”
3. Continue filling pieces into the container (increase the load). Press g/W
and increase the reference sample quantity to the new number of pieces,
or enter the number of pieces manually using the 0 9 numeric keys. Press
u/w to start the counting function. The weighed value for calculation
of the reference weight will now be stored with a longer weighing time and
“increased stability.”
Once the weighed value has been stored, the P symbol remains displayed and
the scale returns to the fast weighing mode.
Monitoring of reference storage accuracy Code Changes
Without** 3 10 1 *
99.0% ( 100 display updates) 3 10 2
99.5% ( 200 display updates) 3 10 3
99.8% ( 500 display updates) 3 10 4
99.9% (1,000 display updates) 3 10 5
** = factory setting
** = minimum load (1display update). If the weight on the scale is lighter, the error
code “Err 22” is displayed.
3–9