Loading...
ISO/GLP Printout
You can have the parameters pertaining
to weighing conditions printed before
(GLP header) and after (GLP footer) the
values from the weighing series.
These parameters include:
GLP header:
Date
Time at the beginning of a weighing
series
Balance manufacturer
Balance model
Model serial number
Software version
ID number
GLP footer:
Date
Time at the end of the weighing series
Field for operator signature
Operating the Balance with an
ISO/GLP-compliant Logging Device
(Printer)
ISO/GLP-compliant documentation
requires a computer with special soft-
ware. Contact Sartorius for a detailed
description.
Setting:
Setup: Printout: ISO/GLP printout:
Always on
The record is output to a Sartorius
YDP03-0CE Data Printer or a computer.
End GLP printout:
Press c
End GLP printout while application
is active:
This requires the following settings:
Setup: Device settings: Keys:
CF function in application:
Clear only selected applications
Press c
> Text line: CF selected: clear application
Press the GLP soft key
--------------------
17.01.2000 16:12
SARTORIUS
Model ME215S
Ser. no. 91205355
Ver. no. 01-41-05
ID 12345678901234
--------------------
L ID 12345678901234
nRef 10 pcs
wRef 1.35274 g
Qnt + 235 pcs
Qnt + 4721 pcs
S ID 12345678901234
Qnt + 567 pcs
--------------------
17.01.2000 16:13
Name:
--------------------
--------------------
17.01.2000 16:24
SARTORIUS
Mod. ME215S
Ser. no. 91205355
Ver. no. 01-41-05
ID
--------------------
LID
Internal calibration
Start: manual
Diff. + 0.06365 g
Internal adjustment
completed
Diff. + 0.00000 g
--------------------
17.01.2000 16:25
Name:
--------------------
Dotted line
Date/time
Balance manufacturer
Balance model
Balance serial number
Software version
(display and control unit)
Balance ID no.
Dotted line
Weighing series no. (lot)
Application initialization value
Application initialization value
Counting result
Counting result
ID for counting result
Counting result
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line
Record of Internal
Calibration/Adjustment:
Dotted line
Date/time
Balance manufacturer
Balance model
Balance serial number
Software version
(display and control unit)
Balance ID no.
Dotted line
Weighing series no. (lot)
Calibration (lot) adjustment mode
Start mode for calibration/adjustment
Difference after calibration/adjustment
Confirmation of completed
calibration/adjustment routine
Difference between current and target val-
ues after calibration
Dotted line
Date/time
Field for operator signature