Internal Calibration/Adjustment
First set either
Internal
cal./adjustment or
Selection mode (factory
setting) in the Setup menu
(Balance/scale functions:
Calibration/adjustment: CAL key
function).
Inside the balance housing are built-in
motorized calibration/adjustment
weights.
The internal calibration/adjustment
sequence is as follows:
Unload the balance
Select the calibration function: Press
the
Cal soft key and then the
Start soft key
> The internal calibration weight
is applied automatically
> The balance is calibrated
> If the setting
Calibrate,
then auto adjust (factory
setting) is selected in the Setup menu,
the balance is now automatically
adjusted
> If the setting
Calibrate,
then manual adjust
is selected in the Setup menu,
the internal calibration routine is now
ended without adjusting the balance
(otherwise, see “Calibration and
Adjustment Sequence,” on next page)
> The internal calibration weight
is removed
> ISO/GLP printout:
see page 131
Internal Linearization
In the Setup menu (Balance/scale func-
tions: Calibration/adjustment: CAL key
function), you need to have set
Internal linearization
or select this using the Selection
mode (factory setting).
Inside the balance housing are built-in
motorized calibration/adjustment
weights for linearization.
The internal linearization sequence
is as follows:
Unload the balance
Select the linearization function: Press
the
Cal soft key, the Select soft
key, and then the
Start soft key
> The internal linearization weights are
applied automatically.
> The balance is linearized
> The internal linearization weights are
removed
> The balance is adjusted automatically
following internal linearization
> ISO/GLP printout: see page 131