24
Commands (Data Input Format Compatible with Current Weigh Cells)
The computer connected via the data port can send commands to the weigh cell for controlling functions. The commands sent are control 
commands and may have different formats. Control commands consist of up to 13 characters. Each character must be transmitted according 
to the settings configured in the operating menu for data transmission. 
Formats for Control Commands
Format 1:  Esc  !  CR  LF
Format 2:  Esc  !  #  _  CR  LF
Esc:  Escape (optional)  CR:  Carriage return
!:  Command character  LF:  Line feed (optional)
_:  Underline
  Command character   Format 1:
  !  Meaning
  K  Ambient conditions: Very stable 
  L  Ambient conditions: calm
  M  Ambient conditions: unstable
  N  Ambient conditions: Very unstable 
  O  Block keys
  P  r key (print, auto print; activate or block)
1
) 
  Q  Acoustic signal
  R  Unblock keys
  S  Restart/self-test
  T  Tare: w key
  U  Tare
  V  Zero
  W  Calibrate | Adjust depending on menu setting
  Z  Perform internal calibration/adjustment
2
)
  Command character   Format 2:
  !#  Meaning
  f0_  Function key 
b
  f1_  function key Q Calibration/adjustment (depending on the menu setting)  
  f2_  Function key u
  s1_  With »s8_« compatibility: Toggle selection in steps of 1 
    With »s9_« compatibility: Adjust according to menu setting
  s2_  Activate parameter mode (selection)
  s3_  c key
  s8_  Compatibility: Consistent with current weigh cells (from 2013)
  s9_  Compatibility: Consistent with older weigh cells (previous models)
  x0_  Perform internal adjustment
  x1_  Print model
  x2_  Print serial no.
  x3_  Print software version
1
) When initiating the print command, the data output rates may differ: see table on next page.
2
) only on models with built-in motorized calibration weight