Operating Instructions YTM15MA 13
Operation
Reasons for an incorrect measurement may be:
− Non-compliance regarding ambient conditions.
− Intense direct sunlight.
− Strong draft.
− "ReproEasy Pad" was not used correctly during the
performance test. E.g. the protective film was not
removed from the "ReproEasy pad" or the incorrect
side of the "ReproEasy pad" was placed on the pan.
− Weighing system is defective.
4.4 Calibrate Heating Module
(MA37 | MA160)
Calibration of the heating module involves calibration
at a fixed temperature setting. The result is saved in
the calibration report.
Carry out the calibration at the subsequent installation
site in order to ensure precise calibration of the
heat ing module. The calibration takes at least
50 minutes.
System Requirements
− At least 30 minutes have passed since the last
application measurement.
− The device has cooled down with the hood open.