22 Operating Instructions YTM15MA
Operation
4.6 Single-Point Temperature
Adjustment (MA35)
Carry out the calibration at the subsequent installation
site in order to ensure precise calibration of the
heating module. The calibration takes around
85 minutes.
System Requirements
− At least 30 minutes have passed since the last
application measurement.
− The device has cooled down with the hood open.
The following procedure corresponds to a fixed,
uniform process between heating up, cooling down,
measuring and adjusting, each under the same general
conditions.
These constant conditions are required so that the
temperature measurements can be compared.
Procedure
t Switch on the device.
t Set the following drying parameters:
− Drying temperature: As set in the method.
− Remaining parameters: Any.
t Select the "Calibrate" function using the cursor
keys.
t Confirm "CAL" (display) with the "ENTER" key.
t Select the "PH" function (heating calibration)
with the cursor keys, and confirm with [ENTER].
y "TAR" appears on the display.
t To tare the device: Press the [Enter] key.