Operating Instructions YTM15MA 9
Operation
The following conditions should be taken into account
when calibrating the moisture analyzer:
Carry out calibration in a temperate area. The room
temperature should be 20°C.
Carry out calibration in a clean measurement
location.
Use a stable and level surface (table).
Drafts, e.g. from open doors/windows, should be
avoided.
If required: Shield the device.
The frequent movement of persons in front of the
device should be avoided (unstable conditions).
Avoid direct sunlight.
Avoid gases or dusts.
Establish a safety distance in the direct vicinity of
the device. At least 20 cm on all sides and at least
1 m above the device.
4.2 Transporting and Removing the
Adjustment Disk
CAUTION
Danger of burns due to hot adjustment disk!
t Always use the disk holder (removal tool) to
transport the adjustment disk or remove it from
the moisture analyzer after calibration.
t Allow the adjustment disk to cool on a clean,
temperature-resistant pad before or after
completion of a measurement.
t Only place the adjustment disk back into the
carrying case after cooling.
t Wear personal protective equipment.