    Angela GouldAug 3, 2025
    What to do if Satelec dental equipment has no spray?
    • D
      David AlvaradoAug 3, 2025
      If your Satelec dental equipment has no spray, several factors could be at play. Start by checking the water supply to ensure the water nose connection isn't defective. Next, verify that there's adequate water pressure from the mains. If those aren't the issue, the filter might be clogged, or the solenoid valve could be defective, requiring you to clean or replace the filter. If problems persist after these checks, return to SATELEC After Sales Service.
  • A
    Ashlee OliverAug 7, 2025
    How to fix Satelec dental equipment with no operation?
    • M
      Martin NelsonAug 7, 2025
      If your Satelec dental equipment isn't operating, there are a couple of things to check. First, make sure the mains cord is properly connected to a working mains socket. Also, check for any moisture, especially on electrical contacts, and dry it thoroughly.
  • L
    Lindsay CurryAug 13, 2025
    Why is there no water at the tip of my Satelec dental equipment but vibration is present?
    • J
      Jason SimmonsAug 13, 2025
      If your Satelec dental equipment has vibration but no water at the tip, it could be due to several reasons. Check if the tip or file is obstructed and remove any obstruction. Ensure that you have selected the correct tip and that the spray is correctly adjusted. Also, a worn or deformed tip can cause this issue, so consider replacing it.
  • S
    Sandra GarzaAug 18, 2025
    How to troubleshoot low vibration in Satelec dental equipment?
    • C
      Collin JohnsonAug 18, 2025
      If you're experiencing low vibration with your Satelec dental equipment, start by cleaning the cord contacts, as a defective connector contact could be the cause. If cleaning the contacts doesn't resolve the issue, the handpiece cord might have a cut wire, in which case you should return it to SATELEC After Sales Service for cord replacement.
  • K
    Kelly BryanAug 22, 2025
    What to do if Satelec dental equipment has insufficient power?
    • R
      Rebecca PierceAug 22, 2025
      If your Satelec dental equipment is showing insufficient power, thoroughly dry the electrical contacts. Also, ensure all connections are tightened with a wrench.
  • B
    Brittney HernandezAug 27, 2025
    How to fix no ultrasonic output in Satelec dental equipment?
    • J
      Jennifer SmithAug 27, 2025
      If there is no ultrasonic output in your Satelec dental equipment, replace the worn sealing gasket (1,15 x 1) in the handpiece with a new gasket (kit F00070).

Summary

Warnings

Installation Procedures

Operating the Device

Maintenance and Sterilization

Handpiece Maintenance and Cleaning

Daily cleaning and rinsing procedures for the handpiece.

Device Daily Maintenance and Inspection

Daily cleaning and inspection of the device and accessories.

Routine Maintenance Tasks

Water Filter Maintenance and Replacement

Procedure for cleaning or replacing the water filter.

Troubleshooting Common Issues

Guide to diagnosing and resolving common device issues.

Satelec SUPRASSON P5 NEWTRON Specifications

General IconGeneral
TypeUltrasonic scaler
Water flow rateAdjustable
Tip VibrationLinear
Operating Frequency28-36 kHz
Power Supply50/60 Hz
Water supplyExternal
Power OutputAdjustable