9 Adjustment
1. switch on the
device
Press the ON/OFF button The green light goes on
2. switch to
calibration mode
Immediately after pressing the
ON/OFF button, press the
PEAK and PRINT buttons
simultaneously, several times in
quick succession, until the
left-hand red light comes on.
The left red light goes on
3. device type Press SET immediately after
the red light comes on.
The maximum Newton
value of the device is
displayed and can now be
set.
3a) (Back in normal
mode???)
(If you have now returned to
normal mode, switch off the
device and start again from
step 1. Press the buttons faster
if necessary)
4. select device Use the ▼▲ buttons to select
the maximum load (N) of the
respective device.
The appropriate value for
the device is shown in the
display.
5. save settings Press SET
6. switch to
calibration
Press MEMORY The right red light comes
on
7. specify the
available calibration
weight
Press UNIT and enter the
calibration weight in Newtons
with ▼▲. (X kg * 9.81)
The weight in newtons is
shown in the display
8. save Press SET and UNIT at the
same time
9. attach weight Hang the weight on the device
and hold it as still as possible.
Then press ZERO
The device switches back
to normal mode and is
calibrated