Monarch Fixed Ceiling Lift Owner’s Manual Part No. 001108-EN, 21-m09-2017
Electromagnetic Immunity
Guidance and Manufacturer’s Declaration -
Electromagnetic Immunity - For All Equipment and Systems
The Monarch Ceiling Lift is intended for use in the electromagnetic environment specified below.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/-8 kV contact
+/-15 kV air
+/-8 kV contact
+/-15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Surge
IEC 61000-4-5
+/-1 kV differential
mode
+/-2 kV for
common mode
+/-1 kV
differential mode
+/-2 kV for
common mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: UT is the AC mains voltage prior to application of the test level.