6. CALIBRATION
6.1 STANDARD CALIBRATION
- Turn off the power, empty the pan, hold the ON/OFF key and the TARE key together until, display
will show 88888, now release the ON/OFF key;
- When the display will shows F1 CAL”, release also the TARE key;
- Press ZERO key to enter setting, display will show UNLDand after the instrument will acquired
automatically the zero value, after that it will show “LD;
-Put the same weight of the maximum capacity on the scale and wait the stability, the display will
show PASS
- Press ZERO key to return in the setup environment.