6.2 LINEAR CALIBRATION
- Turn off the power, empty the pan, hold the ON/OFF key and the TARE key together until, display
will show 88888”, now release the ON/OFF key;
- When the display will shows F1 CAL”, release also the TARE key;
- Press TARE key until the display will show F6 l-c”, at this point press the ZERO key to enter;
- The display will show UNLDand after the instrument will acquired automatically the zero value,
after that it will show “h-CAP”; put, on the scale, half of the capacity and press the ZERO key;
- The display will show “F-CAP”;
- Put the same weight of the maximum capacity on the scale and press the ZERO key, the display
shows PASS”;
- Press ZERO key to return in the setup environment.