KL 1500 compact – Instructions for Use
9/12
4. Maintenance
Your KL 1500 compact is maintenance-free.
There is no provision necessary for disinfecting the light source when using it in a medical field.
To clean the outside of the instrument, use a soft dry cloth or commercially available plastic
cleaning cloths.
5. Troubleshooting
If you are unable to rectify the fault by the actions mentioned as follows, please contact your
specialist dealer or the nearest SCHOTT agency. More extensive repairs must be carried out by
the authorised customer service depot.
Fault Possible causes Remedial action
Lamp out, fan not running Instrument not switched on Switch instrument on
Plug not in socket Plug the plug in
No mains electricity voltage Check mains voltage
Lamp compartment not closed Close lamp compartment
Lampe off, fan running Lamp defective Replace lamp (see point 3 of
these instructions)
No lamp
Lamp no contact
With light source switched off,
install lamp
Push lamp into socket
Electronic overheated Ensure adequate cooling, check
that lamp type is correct, re-start
instrument after cooling down for
a prolonged time
6. Accessories
A wide range of accessories is available for your KL 1500 compact. A separate brochure gives you
comprehensive information (see page 12 for address information).
Only SCHOTT light guides and accessories guarantee perfect operation, safety and optimum light
yield.