PN 10795T 11
B. Press the ON/SELECT button when “CAL” appears. A menu cycle of CAL 1, CAL 2, Attn and
SPd will begin.
C. Press the ON/SELECT button when “CAL 1” appears. The current V/J output sensitivity will be
displayed and the “SET CAL” annunciator will flash.
D. Use the RANGE (count up) button to increase the value or the MODE (count down) to decrease
the value to enter the V/J value listed on the pyroelectric detector’s serial tag.
E. Press the ON/SELECT button to exit this menu. The power meter will return to the last operational
state.
F. Press the MODE button to return to the menu in step A.
G. Press the ON/SELECT button when “CAL” appears. A menu cycle of CAL 1, CAL 2, Attn and
SPd will begin.
H. Press the ON/SELECT button when “SPd” appears. Choose between “hF” for high frequency
absorbers and “bL” for black coated absorbers according to the Table 2 chart.
Table 2. Mode Selection Criteria
Type of Absorber
Pyroelectric Detector Model
Type
Speed Selection
Black Coated
P, SP, PHD and SPHD
bL
High Frequency
PHF and SPHF
hF
I. Press the ON/SELECT button to select the appropriate “SPd” setting. The power meter will return
to the last operational state.
J. Press and hold the OFF/CANCEL button until the display blanks to save the settings.
K. Press the ON/SELECT button to turn the H410 back on.