PN 10795T 12
6. To Select a Range
Note: AUTO range may be selected if the energy levels of repetitive pulses are to be measured. However,
do not select AUTO range if you want to measure single pulse energy or pulses running at repetition rates
lower than 10 Hz.
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the display
blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings are saved, they will
be lost.
A. Press the RANGE button. The analog scale (if active) will disappear and AUTO, 3m, 30m, 300m
and 3 will cycle through the display.
B. To select a range, press the ON/SELECT button when the range you desire appears in the display.
The H410 will return to the last operational state.
C. Press and hold the OFF/CANCEL button until the display blanks to save the setting.
C. Press the ON/SELECT button to turn the H410 back on.
7. To Select a Measurement Mode
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the display
blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings are saved, they will
be lost.
Available measurement modes are:
Energy (J), Avg
Energy (J AVG), and
Power (W)
8. To Measure Energy (J)
Note: The update rate of the display is 20Hz.
The energy mode displays the energy of each pulse of a repetitively pulsed laser or a single pulse.