PN 10795T 22
360401 with interface module 150.0
380401, 380402, 384UV5 with interface module 245.0
360801 with interface module 170.0
380801, 380802, 388UV5 with interface module 280.0
H Press the ON/SELECT button. The speed setting is now active and the power meter will return to
the operational state last used.
I. Press the MODE button. A menu cycle of J, W, CAL, tO and cd will begin.
J. Press the ON/SELECT button when the cd annunciator is displayed. The current calorimeter delay
setting is displayed and the SET annunciator flashes. The calorimeter delay feature prohibits the
display of energy if a pulse is fired before the entered time (1 to 255 seconds) elapses. The
calorimeter must reach environmental thermal equilibrium before a subsequent pulse is fired or
low energy measurements will occur.
To set the time delay between pulses use the count up (RANGE) and count down buttons (MODE)
buttons to enter the time in seconds. Press the SELECT button after the time has been entered to
save the setting. The following time delays are recommended:
AC2500, AC25FX, ACX25FX, ACX25HD 60 seconds
AC2501, AC25HD, AC25UV, AC2504 60 seconds
AC5000, AC50FX, ACX50FX, ACX50HD 90 seconds
AC5001, AC50HD, AC50UV, AC5004 90 seconds
360401 with interface module 105 seconds
380401, 380402, 384UV5 with interface module 180 seconds
360801 with interface module 125 seconds
380801, 380802, 388UV5 with interface module 200 seconds
Note: Make sure the calorimeter delay is less than the time out setting you will make in section
10.
K. Press and hold the OFF/CANCEL button until the display blanks to save the settings.
L. Press the ON/SELECT button to turn the H410 back on.