PN 10795T 23
6. To Select a Range
Note: AUTO range is not available in the energy mode for calorimeters.
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the display
blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings are saved, they will
be lost.
A. Press the RANGE button. The analog scale (if active) will disappear and the available ranges
will cycle through the display. The ranges available for calorimeters are in Table 3.
Table 3. H410 Ranges for Calorimeters
100mm with PN10735
Interface Module*
200mm with PN10747
Interface Module**
Range
100mW
1W
10W
AUTO
100mJ
1J
10J
3W
30W
AUTO
3J
30J
3W
30W
AUTO
3J
30J
3W
30W
AUTO
3J
30J
* With a PN10748 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 50
W or 150J.
** With a PN10769 – 10X attenuator, the actual power or energy is 10 times the displayed value up to
100 W or 300 J.
B. To select a range, press the ON/SELECT button when the range you desire appears in the display.
The selected range will be activated and the H410 will return to the last operational state.
C. Press and hold the OFF/CANCEL button until the display blanks to save the settings.
D. Press the ON/SELECT button to turn the H410 back on.
Note: When using PN10748 or PN10769, 10X attenuators, you may want to use an attenuation factor of
10 so the displayed reading on the H410 is correct. However, the range designation may be incorrect if