PN 10795T 30
Only one detector should be plugged in at any time.
Refer to Figure 1. Slide the CAL/PYRO switch, located on the upper left side of the H410, to the CAL
(up) position. A 3 meter mini DIN type cable comes with Scientech photodiode detectors. Connect the
cable to the power meter’s mini DIN connector.
5. To Configure the H410 for a Photodiode Detector
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the display
blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings are saved, they will
be lost.
To enter into the configuration mode:
A. Press the MODE button. A menu cycle of W, CAL and Attn will begin.
B. Press the ON/SELECT button when W appears. A menu cycle of U and UIS will begin. Choose
U (ultraviolet) for the model AP30UV or UIS (visible) for the model AP30.
C. Press the ON/SELECT button when the appropriate option is in the display. The H410 is now
configured for the chosen photodiode detector and will return to the last operational state.
D. Press and hold the OFF/CANCEL until the display blanks to save the setting.
A. Press the ON/SELECT button to turn the H410 back on.
6. To Select a Wavelength
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the display
blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings are saved, they will
be lost.
To enter into the wavelength selection mode:
A. Press the MODE button. A menu cycle of W, CAL and Attn will begin.