PN 10795T 31
B. Press the ON/SELECT button when CAL appears. The current wavelength is displayed and the
SET annunciator is flashing.
D. Use the RANGE (count up) and MODE (count down) buttons to change the display (in 1 nm
increments) to the desired wavelength. The available wavelengths are as follows:
200 nm to 1100 nm for Model AP30UV
400 nm to 1100 nm for Model AP30
D. Press the ON/SELECT button. The new wavelength setting is now active and the H410 will return
to the last operational state.
E. Press and hold the OFF/CANCEL until the display blanks to save the setting.
B. Press the ON/SELECT button to turn the H410 back on.
7. To Select a Range
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the display
blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings are saved, they will
be lost.
A. Press the RANGE button. The analog scale (if active) will disappear and AUTO, 30μW, 300 μW,
3mW and 30mW will cycle through the display.
B. Press the ON/SELECT button when the range you desire appears in the display. The selected
range is activated and the H410 will return to the last operational state.
C. Press and hold the OFF/CANCEL button until the display blanks to save the setting.
D. Press the ON/SELECT button to turn the H410 back on.
8. To Zero the Display
A. Press the CANCEL button before the photodiode is lased.