14
11.1.2 Available rotors and adapters
Rotor
type
Tubes
Tubes
Per
Rotor
Dimension
(Ф×L
mm)
Adapters
Maximum
speedrpm
Radius
cm
Maximum
RCF
(×g
A12-10P
15ml con 6 17×120 4500 11 2490
1.5-5ml vacu 12 13×82
A10P15
mat
4500 9.8 2218
4-7 ml vacu 12
13×106
A10P15
4500
11 2490
16×75
A10P15
Mat
9.8 2218
8.5-10 ml vacu 12 16×107 A10P15 4500 11 2490
2.7-3
(EU) ml
collection tube
12 11×66
A10P15
mat
4500 9.8 2218
7.5-8.2
(EU) ml
collection tube
12 15×92 A10P15 4500 11 2490
Table 11.1 Rotors and adapters
11.1.3 Notice
z The centrifuge rotor can separate samples with a density lower than 2.0g/ml. If the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
11.1.4 Autoclaving
A12-10P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.